SELECTABLE TEMPERATURE BETWEEN 70 ºC & 121ºC.
CAPACITY: 20, 80 & 150 L.
AUTOCLAVES “MAN” MODEL NEED A COMPRESSED AIR EXTERNAL INSTALATION.
AUTOCLAVES “COM” MODEL INCLUDE COMPRESOR AND DO NOT NEED COMPRESSED AIR INSTALLATION.
WATER SUPPLY NEEDED WITH LOW-CAL CONTENT AND DRAINAGE. (WATER PRESURE: 3 BAR MINIMUM).
CONFORMITY
EN 61010-1 SAFETY FOR EQUIPMENTS FOR LABORATORY USE .
EN 61010-2-040 SAFETY FOR STERILIZER EQUIPMENTS.
EN 61326 ELECTROMAGNETIC COMPATIBILITY.
EN/97/23/EC SAFETY FOR PRESSURE EQUIPMENTS.
APPLICATIONS
Preserved food sterilization.
Very suitable for small productions and wide variety.
FEATURES
External case, top, tank and lid made of stainless steel.
Tank fill by external water inlet.
Temperature control probe inside the sample.
Counter pressure system to prevent the cans breakage.
Temperature, time and pressure registered in USB.
SAFETY
Device that avoids opening the lid when the chamber is under pressure.
Safety valve:Prevents overpressure over the maximum limit.
Safety pressure switch:Disconnects heater when there’s overpressure.
Thermally insulatedlid. Doorproperly closed detector.
Doublefilter retention
CONTROL PANEL
1. Parameters LCD display.
2. Microprocessor single button.
3. Operation Led.
4.Start/stop push button.
5. USB adapter.
6. Open lid knob.
7. USB function light indicator.
Model, Kat.br. | Drying | Capacity litres | Ø Heigh (usable) cm | Height/Width/Dept (exterior) cm | Temp, °C | Capacity baskets | Power W | Weight Kg |
---|---|---|---|---|---|---|---|---|
STERIL-FOOD CAM 20, 4002421 | NO | 20 | Ø24, 41cm | 70x38x38 | 70 to 121 | 2 | 2500 | 55 |
STERIL-FOOD MAN 80 | NO | 80 | Ø40, 60cm | 107x58x72 | 70 to 121 | 2 | 3600 | 136 |
STERIL-FOOD COM 80 | NO | 80 | Ø40, 60cm | 107x58x72 | 70 to 121 | 2 | 3600 | 136 |
STERIL-FOOD MAN 150 | NO | 150 | Ø50, 70cm | 118x80x95 | 70 to 121 | 3 | 7500 | 250 |